Conducting a Wipe Test

Purpose
All Principal Investigators actively using radioactive material must insure
that routine wipe tests are conducted in their laboratories. Normally these wipe
tests are conducted on a monthly basis at a minimum. However, if researchers use
large amounts of radioactive material in a laboratory, it may be necessary to
conduct more frequent wipe tests. See the Albert Einstein College of Medicine
Radiation Safety Manual for frequency of wipe based on amount of radioactive
material.
Demonstration (Video):
Conducting a Wipe Test
Definitions
- Background Sample (Blank): A clean cotton swab or filter
paper used to determine the number of counts corresponding to a
contamination free surface.
- Control Sample: A scintillation vial containing a known
amount of radioactive material, typically less than 0.1 microCurie, and a
clean cotton swab or filter paper.
- Disintegration per minute (dpm): the number transformation
from the radioisotope to its progeny isotope in a one minute period. For
example, P-32 decays to S-32 at a certain rate that can be described in dpm.
- Inactive Status: Principal Investigators who are not using
radioactive material for an extended period may choose to place their
license on inactive status. When the license is on inactive status, the
Principal Investigator does not need to perform wipe tests in his/her
laboratory.
Procedure
A map of the lab should be created showing benches, refrigerators, freezers,
sinks, fume hoods, and biosafety cabinets. Equipment used in radioactive research
such as centrifuges, hybridization ovens and water baths should be drawn on the map
as well. The map should
indicate areas to be checked for contamination using the wipe test method. Use
numbers to correspond to the various locations. The
number of wipes to be conducted depends on the size of the laboratory and the
number of laboratory benches used for radioactive research. For example, if two
benches are used for radioactive research than those two benches should be
monitored for contamination each month. Here are some typical locations that
should be monitored for contamination using the wipe test method:
- Laboratory benches used for radioactive material research
- Floors in front of benches
- Equipment used with radioactive material
- Door handle to freezer and/or refrigerator used for storing radioactive
material
- Sink used for disposal of radioactive material
- Cover to radioactive waste containers
- Fume Hoods used for radioactive material research
- Laboratory door handle
In addition, you must monitor equipment used in radioactive research,
including:
- Centrifuges
- Hybridization ovens
- Water baths
- Scintillation counters and/or gamma counters
- Plexiglas shields
A cotton swab or filter paper may be used to wipe the surface of the target
area. Moisten the swab/filter paper and wipe approximately 100 cm2.
Perform the wipe test on all designated areas on the map using a new swab/filter
paper for each location. When analyzing the wipes, a background sample and
control must be included. The background and control are used to determine if a
particular wipe is contaminated and the amount of contamination.
Place the swabs/filter papers into scintillation vials and add the
scintillation fluid. Cap the vials and record the number for that wipe on the
cap. Place the vials into a scintillation rack and place the rack into the
scintillation counter. Set the scintillation counter to detect the radioisotopes
used in the laboratory and run the samples. The results of the analysis is
typically printed out on paper. It displays the sample number, the time, the
counts and the count rate in cpm.
As a rule-of-thumb if any of your sample results are 3 to 4 times background,
it is considered contaminated. For example if your background sample is 10 cpm
and you have a wipe from your lab bench that is 50 cpm, than the bench is
contaminated. If you find that one of the locations is contaminated you must
determine the amount of contamination on the wipe using the background and
control results. To do this you may use the
Calibration Work Sheet, inserting
the numbers from the results of analysis in the liquid scintillation counter.
Locate the surface that is contaminated and take the necessary steps to
decontaminate the area in accordance with Section 20.0 of the Radiation Safety
Manual titled "Emergency Response".
Recordkeeping
The results of wipe tests must be kept in the Radiation Safety Manual for a
minimum of five years. The records must contain the laboratory map, the results
from the scintillation counter and Calibration Work Sheets for contaminations.
Radiation Safety |